Literature for Humerus Interlocking
Features
1. Antegrade as well as rectograde nailing
2. Distal Locking anteroposterior as well as Lat-medical
3. May eliminate an existing fracture gap of 10mm
4. Solid Nail. Subsequent Compression, May also be given after osteolysis at the fracture site
5. Available in different length varying at 1cm from 19cm to 31cm
6. The end cap (Top Screw) protect the inner thread of the nail from the tissue in growth and facilitates subsequent implant removal. End caps are provided in 4 lengths. 0,5,10,15mm extension, which helps to extend the nail length if desired. This allows free positioning of the locking holes into areas with good bone quality.
Provision of oblique placement of screw in proximal fragment.
UHN Ordering Information
1 Insertion Handle for UHN
2 Compression Device for UHN
3 Connecting Screw for UHN
4 Compression Connecting Screw
5 Coupling Block for UHN extraction
6 Inserter / Extractor for UHN
7 Slotted Hammer
8 Awl with T-Handle (Cannulated)
9 Projectile Reamer (Coupling end)
10 Protection Sleeve 11.0 / 8.0
11 Drill Sleeve 8.0 / 2.7
12 Trocar 8.0mm
13 Depth Guage for Locking bolt
14 Drill Bit 2.7mm Extra Long (Q.C)
15 Combination Wrench
16 Distal Jig
Antegrade Insertion
Humeral shaft fractures, distal and proximal
third
Patient positioning : Place
the patient supine in a semi-reclined position with a bolster under the shoulder.
Turn the patients head away from the injured side to maximize exposure
of the shoulder
Determinattion of implant Length:- Determine
the approximate nail length preoperatively by measuring the uninjured humerus
from the top of the humeral head to the superior aspect of the olecranon fossa
and subtracting 3-4 cmThe correct implant length can only be determined at
the injured limb after fracture reduction
Insertion site preparation:
The entry point must be in line with the medullary
canal. This is usually at the margin of the articular surface medial
to the greater tuberosity.Incisr the skin just lateral to the acromion, and
split the deltoid muscle in line with its fibres. Palpate the tuberosity and
identify the supraspinatus tendon, without exposing it. Incise the mid-tendon
in line with its fibres. Be careful not to injure the rotator cuff. The arm
can be adducted across the chest to gain better access to the proximal humerus
Using the straight Cannulated Awl with T-handle, insert the K-Wire into the
proximal humerus at the correct entry point. Pass the wire under image intensification
directly into the medullary canal, release the wire by loosening the nut,
and open the medullary cavity eith the awl
Nail/ Instrument Assembly : Couple
the correct length nail with the insertion handle making sure that the apex
of the nails bend is pointing away from the insertion handle. Manually
thread the connecting screw into the nail, and tighten with the 11mm Combination
Wrench
Important:- If there is a fracture gap to eliminate,
the compression device has to be coupled to the insertion handle..
Nail Insertion:- Insert
the nail manually as far as possible. Continue to insert the nail up to the
fracture line, then reduce the fracture and monitor the passing of the fracture
line with the image intensifier. Manipulate with care especially in fractures
in the range between the middle to distal third in order not to damage the
Nervus radialis
ncase of preoperative radial nerve palsy, an exploration of
the radial nerve through a short anteriolateral incision at the transition
of the middle to the distal shaft third may be necessary
If needed, the Insertior / Extractor can be threaded onto the end of the connecting
screw. Light blows with the slotted hammer may support the insertion but are
normally not needed. Do not tap on the insertion handle, this might impair
its accuracy

If the nail does not pass easily, it may be adequate
to enlarge the humeral opening with a hand reamer in order to avoid latrogenic
fractures. Never insert the nail under great force
Pressure against the humeral head when inserting the nail prevents diastasis
and disturbances in the healing process possibly related to it. Bury the nail
completely into the humeral head preferably to 10mm to avoid subsequent irritation
of the shoulder structure even when the arm is abducted (impingement syndrome).
Finally nail length may be adjusted with the end cap.
Distal Locking:-
Reconfirm reduction of the distal
fragment as inserting the nail may create a fracture line.
If the nail tip has reached its final position, first lock distally with the
distal locking device or with standard freehand locking technique. Be careful
not to injure the brachial artery or the median nerve with the drill bit.
Determine the correct bolt length with the Depth guage for Locking bolt
Proximal Locking:-
Lock proximally by means of the insertion handle.If you
use the caliberated drill bit, you may directly read the correct locking bolt
length off the end of the end of the drill bit protruding out of the drill
sleeve.
Alternatively use the Depth guage for Locking Bolts.
Note:- When using the depth guage for locking bolts, add 2 mm to the measurement
to ensure full engagement of the far cortex.
In each main fragment, especially in short ones,
place two locking bolts.
lease note that the placement of a locking bolt in the oblique locking hole
prohibits placement of a second locking bolt through the transverse hole or
through the slot.
Compression Device:-
The humerus
is a hanging extremity. As opposed to femur and tibia when dynamically loaded
there is not automatically a compact reduction. A fracture gap can lead to
a non-or delayed-union, though.
The compression device allows a sensitive compression of the two fragments
in order to eliminate an existing fracture gap.
The compression device is fixed to the nail with the compression connecting
screw. The nail is inserted into the medullary canal and locked at the tip.
A bolt is introduced into the compression slot. By tightening the nut, the
bolt and with it the whole fragment are moved towards the nail tip. Check
under the image intensifier when the gap is closed.
It is important to take into account the protruding of the nail at the insertion
site after the compression.
To hold the achieved reduction, a second bolt has to be introduced into the
static hole. The compression device is now removed and an end cap is inserted
into the nail.
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
||
![]() |
||||||||
|
|
||||||||
|
|
||||||||
|
|
||||||||
|
|
|